Model Number CYF-5 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has been returned to olympus, however, the device evaluation has not been completed.Upon completion of the investigation, a supplemental report will be submitted.
|
|
Event Description
|
A user facility reported to olympus that while removing the stopcock for cleaning, the entire port came out of the scope.The problem, as reported to olympus, was identified during reprocessing.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the legal manufacturer's investigation, it is likely there was an external force applied to the a-rubber, which caused it to become dislodged.In addition, the t-tube mounting cap (instrument channel port) was also subject to external force when attaching or detaching the t-tube (forceps / irrigation plug), which likely resulted in damage (loosening and/or deformation).It is also confirmed that the device was manufactured more than 4 years ago, which use over time could lead to aging deterioration.
|
|