Model Number CYF-5
Device Problem Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The suspect device has been returned to olympus, however, the device evaluation has not been completed.Upon completion of the investigation, a supplemental report will be submitted.
 
Event Description
A user facility reported to olympus that while removing the stopcock for cleaning, the entire port came out of the scope.The problem, as reported to olympus, was identified during reprocessing.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the legal manufacturer's investigation, it is likely there was an external force applied to the a-rubber, which caused it to become dislodged.In addition, the t-tube mounting cap (instrument channel port) was also subject to external force when attaching or detaching the t-tube (forceps / irrigation plug), which likely resulted in damage (loosening and/or deformation).It is also confirmed that the device was manufactured more than 4 years ago, which use over time could lead to aging deterioration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11345734
MDR Text Key232587782
Report Number8010047-2021-02845
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/26/2021
Initial Date FDA Received02/18/2021
Supplement Dates Manufacturer Received03/09/2021
Supplement Dates FDA Received03/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-