Model Number V60 PLUS
Device Problems Break (1069); Failure to Zero (1683)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Date of report: 18feb2021.
 
Event Description
It was reported there was a proximal pressure sensor auto zero failure with alarm at the power on self test.The ventilator was in use at the time of the issue and no patient harm was reported.The proximal pressure sensor auto zero failure was confirmed in the error log and it was noted there was a chip in the touchscreen.Solenoid pin were inspected and were ok.The engineer replaced the data acquisition board and the touchscreen.The issue was resolved.
 
Manufacturer Narrative
The data acquisition board was received for failure investigation(fi).No-fault found.Fi investigation could not replicate the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11346390
MDR Text Key240468692
Report Number2031642-2021-00658
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838089280
UDI-Public(01)00884838089280
Combination Product (y/n)N
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60 PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/18/2021
Supplement Dates Manufacturer Received09/12/2021
Supplement Dates FDA Received10/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment