Additional information was added to h3, h4 and h6.H4: the lot was manufactured from september 01, 2020 - september 02, 2020.H10: the device was received containing 99 ml of fluid in the bladder.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.After the luer cap was removed, evidence of continuous flow of fluid was observed coming out at the distal luer.A functional flow rate test was performed and the results were within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|