• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4702K
Device Problem No Flow (2991)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
It was reported that the event occurred on an unknown day in (b)(6) of 2021.Initial reporter postal code: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume folfusor did not flow.This occurred prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from september 01, 2020 - september 02, 2020.H10: the device was received containing 99 ml of fluid in the bladder.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.After the luer cap was removed, evidence of continuous flow of fluid was observed coming out at the distal luer.A functional flow rate test was performed and the results were within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11350780
MDR Text Key232835152
Report Number1416980-2021-00755
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080154
UDI-Public(01)00085412080154
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4702K
Device Lot Number20J030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2021
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/19/2021
Supplement Dates Manufacturer Received02/21/2021
Supplement Dates FDA Received03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-