Catalog Number A100400000000 |
Device Problems
Detachment of Device or Device Component (2907); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported that a x-smart contra angle won't hold files; no injury resulted.
|
|
Manufacturer Narrative
|
X-smart head cap.Bad maintenance (user).There is some rust on the head cap.X-smart cartridge.Various mechanical problem.The file does not fit on the contra angle because hook of cartridge is broken.X-smart neck.Other.Rotation not fluid.Replaced 1 x x head cap h1004c5210500, 1 x x cartridge h1004c5210150 and 1 x x neck h1004c5470220.
|
|