• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4009K
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor leaked.The device was filled with 3500mg 5-fluoruracil in 230ml 0.9% sodium chloride.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photograph was performed using the naked eye which did not identify any abnormalities that could have contributed to the reported condition; the device was identified in a yellow bag with no evidence of a leak.Due to the nature of the sample, no additional testing could be performed.The reported condition could not be verified through photo inspection.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11353721
MDR Text Key233031916
Report Number1416980-2021-00774
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080130
UDI-Public(01)00085412080130
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4009K
Device Lot Number20J025
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/19/2021
Supplement Dates Manufacturer Received03/04/2021
Supplement Dates FDA Received03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-