Model Number 26000009
Device Problems Separation Failure (2547); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Unequal Limb Length (4534)
Event Type  Injury  
Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event Description
Allegedly, the patient was revised due to leg length discrepancy of about 4mm on the right side.Radiographic images showed possible loosening of cup and stem.The revision was scheduled to replace the head and neck and possibly the liner.When trying to remove the neck from the stem, it could not be separated.So the stem was also removed.Additional information received on 02/02/2021: revised part and lot numbers were provided as well as the reason for revision, moreover the revision surgery date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERAMIC FEMORAL HEAD
Type of Device
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key11357861
MDR Text Key232808612
Report Number3010536692-2021-00096
Device Sequence Number1
Product Code MRA
UDI-Device IdentifierM684260000091
UDI-PublicM684260000091
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model Number26000009
Device Catalogue Number26000009
Device Lot Number094113340
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/02/2021
Initial Date Manufacturer Received 02/02/2021
Initial Date FDA Received02/22/2021
Supplement Dates Manufacturer Received02/02/2021
Supplement Dates FDA Received04/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-