• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9736242
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9736226, serial/lot #: (b)(4).A medtronic representative went to the site to test the equipment.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system that was used outside of procedure.It was reported that site was trying to load images via cd and one disc worked and one disc did not work.The disc that did not work had the system indicate that it did not have images that could be found.Other navigation systems can see dicom images on the same disc.No patient present.
 
Manufacturer Narrative
H3: the software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.H6: conclusion code d14 and results code c19 apply to the system.Conclusion code d02 and results code c10 apply to the software.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION FLEXENT
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11359358
MDR Text Key232872917
Report Number1723170-2021-00460
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
PMA/PMN Number
K200723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9736242
Device Catalogue Number9736242
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/18/2021
Initial Date FDA Received02/22/2021
Supplement Dates Manufacturer Received05/20/2021
Supplement Dates FDA Received05/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."
-
-