Model Number 37612 |
Device Problems
Unstable (1667); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Charging Problem (2892); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 37651, serial#: (b)(4), product type: recharger.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received by a manufacturing representative (rep) regarding a patient who was implanted with an implantable neurostim ulator (ins).It was reported that the patient was having issues charging.They had fluctuating coupling bars.The patient's nursing home staff let the device deplete and the rep had to do a prm.As they were charging with the rep, they noticed more fluctuating co upling bars.Rep said the ins may be tilted a little.The patient wants a wireless recharger.The old recharger is being sent back.No symptoms were reported.
|
|
Event Description
|
Additional information was received by the manufacturer representative that it looks as if the patients device in implanted at an slight angle.Cause is unknown.Coupling issues have resolved and charging is much better.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 37651; serial# (b)(6).Product type recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|