• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the perforator plunged.No additional information was provided.
 
Event Description
N/a.
 
Manufacturer Narrative
The perforator was not returned for evaluation (discarded by customer); therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA 02048
MDR Report Key11367831
MDR Text Key249448055
Report Number3014334038-2021-00028
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/01/2021
Initial Date FDA Received02/23/2021
Supplement Dates Manufacturer Received03/05/2021
Supplement Dates FDA Received03/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-