Model Number VICMO12.6
Device Problems Crack (1135); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2020
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.Work order search: no similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon removed a 12.6mm vicmo12.6 implantable collamer lens, -05.00 diopter, from the bottle and noted there were cracks on the surface of the lens.There was no patient contact.Another same model/length lens was implanted on (b)(6) 2020 and the problem was resolved.Cause of the event was unknown.
 
Manufacturer Narrative
H3: device evaluation: lens was returned inside a micro centrifuge vial with moisture on the lens.Visual inspection found the haptic torn.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key11368713
MDR Text Key233987509
Report Number2023826-2021-00465
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
PMA/PMN Number
NA#S-P#NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Model NumberVICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received02/23/2021
Supplement Dates Manufacturer Received08/16/2021
Supplement Dates FDA Received08/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age28 YR
-
-