According to the available information, the device was checked for integrity and nothing was visible to the naked eye.The catheter was inserted on (b)(6) 2021.Three pieces of green debris were found in the patient¿s urinary tract, reported to be green debris from the guide (fibers in the order of millimeters or even micrometers).An additional uteroscopy was necessary to remove the debris.The guide was checked for integrity after its ¿release¿; there was no breakage and no evidence to explain the green debris in the urinary tract.The guide was also not folded.
|
Visual inspection of the returned guidewire found tiny holes in the coating, and three soft bends: one approximately 20cm from the distal end, one approximately 60cm from the distal end, and one approximately 110cm from the distal end.Inspection with a microscope revealed severe scratch marks on the wire, worst in the first 20cm from the distal end and in the last 50cm before the proximal end.Along the guide wire are also mnior scratch marks.The scratches go in both directions, indicating that the guide wire had been pulled back and forth over a metal device.The scratched coating material has been pushed, but because the material is flexible, most of it is still attached.The surface of the coating is smooth and uniform where it has not been scratched and there are no signs of flaking.The damage afflicted to the guidewire indicates that the guide wire had been in interaction with a hard device (metal or likewise hard).Lot review found the device met specifications prior to release.
|