Manufacturer's investigation conclusion: the reported event of a kink in the centrimag motor cable was confirmed via evaluation of the returned centrimag motor (serial number: (b)(4)), as a kink was observed on its cable near the motor end bend relief upon arrival.The motor was tested alongside known working test centrimag equipment and was found to perform as intended despite the observed damage.The motor successfully operated in a mock loop without any issues, including when the motor cable was manipulated by hand.The motor was scrapped as a result of the observed irreparable cable damage.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed for the centrimag motor (serial number: (b)(4)) and the records revealed that the centrimag motor was manufactured in accordance with manufacturing and qa specifications.The centrimag motor was shipped to the customer on 12jun2020.The 2nd generation centrimag system operating manual section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.Centrimag motor instructions for use instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
|