• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH RENTAL, CENTRIMAG MOTOR; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH RENTAL, CENTRIMAG MOTOR; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number L102956
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2020
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a kink in the centrimag motor cable was confirmed via evaluation of the returned centrimag motor (serial number: (b)(4)), as a kink was observed on its cable near the motor end bend relief upon arrival.The motor was tested alongside known working test centrimag equipment and was found to perform as intended despite the observed damage.The motor successfully operated in a mock loop without any issues, including when the motor cable was manipulated by hand.The motor was scrapped as a result of the observed irreparable cable damage.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed for the centrimag motor (serial number: (b)(4)) and the records revealed that the centrimag motor was manufactured in accordance with manufacturing and qa specifications.The centrimag motor was shipped to the customer on 12jun2020.The 2nd generation centrimag system operating manual section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.Centrimag motor instructions for use instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
 
Event Description
It was reported that the motor was returned for analysis.Upon inspection it was found to have a kink in the cable at the motor end bend relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENTAL, CENTRIMAG MOTOR
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key11377568
MDR Text Key234224834
Report Number3003306248-2021-01064
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public07640135140078
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberL102956
Device Catalogue Number102956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/22/2021
Initial Date FDA Received02/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberFA-Q318-MCS-1
Patient Sequence Number1
-
-