• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number FLOW-I C20
Device Problem Output Problem (3005)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2021
Event Type  malfunction  
Event Description
It was reported that the anesthesia workstation failed the safety valve test during system checkout.There was no patient involvement.Manufacturer's reference #: (b)(4).
 
Event Description
Manufacturer's reference #: (b)(4).
 
Manufacturer Narrative
No investigation of the anesthesia system has been performed.The customer has not accepted the service offer.No parts or device logs have been available.With this limited information the root cause to the reported issue can not be determined.During the safety valve test the opening pressure of the safety valve is calibrated.The safety valve protects the patient from high pressures.H3 other text : 4117.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key11377937
MDR Text Key233463378
Report Number8010042-2021-00496
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K191027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLOW-I C20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received11/30/2022
Supplement Dates FDA Received12/07/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-