• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 81000
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site and was able to confirm the reported condition.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Root cause: the root cause of the iv pole easily falling was the adjustment of the iv pole latch.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
3032314970
MDR Report Key11379020
MDR Text Key262227688
Report Number1722028-2021-00083
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK170059
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 02/04/2021
Initial Date FDA Received02/25/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-