• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. NRFIT; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. NRFIT; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 21-7609-24
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Customer did not provide the lot number of the affected device.
 
Event Description
Information was received regarding a cadd cassette reservoir.It was reported that immediately after starting to use the product, the customer noticed medical fluid was leaking from the connector.There was no patient injury.
 
Manufacturer Narrative
Returned device was received for evaluation.During the evaluation of the device, the customer reported condition was confirmed.Problem source was traced to manufacturing.The device history record was reviewed and showed, that this device met all manufacturing specification for product released for distribution.No issues were identified, that would have impacted this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NRFIT
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key11381943
MDR Text Key233732442
Report Number3012307300-2021-01578
Device Sequence Number1
Product Code LHI
UDI-Device Identifier10610586044014
UDI-Public10610586044014
Combination Product (y/n)N
PMA/PMN Number
K162219
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number21-7609-24
Device Catalogue Number21-7609-24
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/01/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received03/31/2021
Supplement Dates FDA Received04/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-