• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 DF1 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 DF1 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1231-40
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Patient Involvement (2645)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Event Description
This report is to advise of an event observed during analysis.The implantable cardioverter defibrillator exhibited premature battery depletion and a battery performance alert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY VR, U1.6 DF1 US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11382685
MDR Text Key233741993
Report Number2017865-2021-09534
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734504355
UDI-Public05414734504355
Combination Product (y/n)N
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2011
Device Model NumberCD1231-40
Device Catalogue NumberCD1231-40
Device Lot Number3123340
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2017
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/23/2021
Initial Date FDA Received02/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/03/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0003-2018;Z-0115-2017
Patient Sequence Number1
-
-