Model Number 867030
Device Problem No Audible Alarm (1019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/08/2021
Event Type  malfunction  
Event Description
The customer reported a speaker malfunction.It is unknown if the device was in use monitoring a patient at the time of the event.No adverse event involving patient or user was reported.The affected intellivue multi meaurement server x3 was sent to the philips bench for further investigation and repair.The reported issue was confirmed by the bench technician.During the investigation of the intellivue multi measurement server x3 the bench technician found the speaker assembly was defective and needed to be replaced.There was no sound coming from the device.The bench technician exchanged the speaker, and the device returned to full functionality.All functional and safety tests were passed, and the device was returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE X3
Type of Device
INTELLIVUE X3
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
derek sammarco
222 jacobs street
cambridge, MA 02141
MDR Report Key11385464
MDR Text Key233798827
Report Number9610816-2021-10006
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838082588
UDI-Public00884838082588
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number867030
Device Catalogue Number867030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2021
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date02/08/2021
Initial Date Manufacturer Received 02/08/2021
Initial Date FDA Received02/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-