Product analysis was completed for the returned device.After review of received data from explanted devices, it was found that high impedance was seen at a much earlier date.Product analysis of the lead was completed.Continuity checks of the returned lead portion were performed during the functional analysis, and no discontinuities were identified.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.Two sets of set setscrew marks were observed on the connector pin.The marks provide evidence of a proper mechanical contact between conductive surfaces of both the generator and connector pin, thereby ensuring a good electrical connection to the lead at one point in time.Abrasions were observed in various areas, likely due to wear.The lead assembly has dried remnants of what appear to have once been body fluids inside the inner and the outer silicone tubing, in some areas.No obvious point of entrance was noted other than the identified cut tube opening and the cut end of the returned lead portion.Note that since a portion of the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Product analysis of the generator was completed.A comprehensive automated pcba electrical evaluation and a battery life calculation were performed.Other than the ifi ¿vboost¿ compliance voltage event, no anomalies were seen, and the device performed according to functional specifications.No other relevant information has been received to date.
|