The suspect device was evaluated and repaired by olympus.The reported insufflation problem was confirmed; the front panel of the insufflation unit was found broken at the insufflation port.When the unit was powered, the 2 green leds of the bar graph pressure indicator lit up, despite no co2 gas inside the unit.The cause of the problem was isolated to a faulty main board.
|
A user facility returned the olympus, uhi-3, high flow insufflation unit for repair for a reported problem with the insufflation port.Upon evaluation of the returned device, it was noted that no co2 gas was supplied and the 2 green led's were lit.There was no patient injury or harm, associated with the problem, reported to olympus.
|