RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 01mar2021.
|
|
Event Description
|
The customer reported resonance was emanated from the inside of the device when the circuit was blocked.The unit was not in use, and there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
G4:16feb2021, b4:10mar2021.The field service engineer (fse) confirmed the reported failure.The fse replaced the blower to resolve the issue.The unit was tested and it was returned to service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4:23apr2021.B4:26apr2021.D4: udi: (b)(4).The blower was returned to manufacturer to failure investigation (fi) for analysis.The blower assembly passed all testing.The reported complaint of a blower assembly noisy was not duplicated.There were no faults found with the blower assembly.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
It was reported that the device was in use on a patient when the failure occurred.No patient information was disclosed by the customer.
|
|