It was reported that during the surgery the pch bskt nrwln upbtr is not cutting properly, hhi instrument is bent from the shaft.No patient injuries or significant delay was reported.Surgery was completed using a competitor device.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The markings on the device confirm the product identification information, and the device is worn from use.There is an upward bent in the shaft.A functional evaluation of the returned device found that it actuates with some resistance.The device is able to cut cleanly through the testing material.A review of the customer provided image found labelling confirming the product identification information.No damage can be seen on the device in the image.The complaint of material deformation was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include wear from repeated use of the device over time and inadequate routine maintenance of the device.The complaint of dull/blunt was not confirmed, and the root cause could not be determined.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|