SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problems
Display or Visual Feedback Problem (1184); Protective Measures Problem (3015); Intermittent Energy Output (4025)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issue.A functional evaluation revealed a hand piece sensor fault.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The complaint was confirmed and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event include a failure of the reed switch or damage on the hall board which may contribute to a short circuit.No containment or corrective actions are recommended at this time.Reference number: (b)(4).
|
|
Event Description
|
It was reported that the motor drive unit was found to have an intermittent power issue, it was cutting in and out.No case reported; therefore, there was no patient involvement.Results of investigation have concluded that the devices revealed a "handpiece sensor fault" during the functional evaluation, which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|