SMITH & NEPHEW MEDICAL LTD. VERSAJET EXACT ASSY, 45 DEGREE X 14MM; LAVAGE, JET Back to Search Results
Model Number 66800041
Device Problems Material Rupture (1546); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/30/2020
Event Type  malfunction  
Manufacturer Narrative
Our reference number: (b)(4).
 
Event Description
It was reported that during the use of the versajet, a patient was undergoing an excision for 3rd degree burns and disassociation of the hose from the water flow at the lock (sterile part) was observed.The problem occurred after 30 minutes of use.The intervention was completed with a new versajet kit.There were no consequences for the patient.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A relationship between the report event and device could be established.A visual inspection was performed and showed the feed line hose was received detached from the pump cartridge.There did not appear to be any residual adhesive within the feed line where it would connect to the pump cartridge fitting.Root cause is determine to be a manufacturing process error.This failure is considered to be isolated.Functional inspection was performed and showed the feed line was connected to the feed line fitting within the pump cartridge.The unit then primed and cut.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET EXACT ASSY, 45 DEGREE X 14MM
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11401836
MDR Text Key234394446
Report Number8043484-2021-00450
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00040565124681
UDI-Public00040565124681
Combination Product (y/n)N
PMA/PMN Number
K110958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66800041
Device Catalogue Number66800041
Device Lot Number50836959
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/05/2021
Initial Date FDA Received03/02/2021
Supplement Dates Manufacturer Received06/08/2021
Supplement Dates FDA Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-