Model Number 4000 |
Device Problem
Device Alarm System (1012)
|
Patient Problems
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation completed on a smiths medical syringe infusion pumps|medfusion 4000 pumps motor rate error was not revealed in the event log and during testing the event was not duplicated as multiple occlusion testing was done.The physical condition of the device reveled a top case chipped and front corner chipped.Preventative action was taken to replace the motor assembly as this was five years old.Device then passed all functional testing.
|
|
Event Description
|
Information received a smiths medical syringe infusion pumps|medfusion 4000 revealed motor error.No patient adverse events reported.
|
|
Manufacturer Narrative
|
Other text: additional information was received by smiths medical on (b)(6) 2021 that the reported problem was found during annual pm testing.
|
|