• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3R
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has been evaluated by olympus.Preliminary findings are reported.The investigation is ongoing.The definitive cause of the user¿s experience cannot be determined at this time.Physical evaluation of the complaint device reveals: there was a leak found at the biopsy channel due to a cut/crack in the rubber glue.The forceps channel was found to be leaking with a scrape mark inside the boroscope.The angulation was low.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event Description
It is reporting during reprocessing of an uretero-reno videoscope using a sterrad reprocessor and a wire/nylon bristle brush for ureteroscopes, after cleaning it several times, black residue keeps coming out of it.The was no patient contact/impact related to this occurrence.
 
Manufacturer Narrative
This report is being updated to provide investigation findings.New information is reported in h4, h6, and h10.The device history record (dhr) for the complaint device has been reviewed and it is confirmed that the device met all design and quality specification when it was shipped.The instructions for use (ifu) shipped with the device provides the user the following information related to the reported event: inspection of the endoscopic system - inspection of the irrigation function precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Conclusion: the definitive cause of the reported events could not be established.Based on the available information, it is possible the internal biopsy channel was scraped forcibly with the insertion of the endo therapy accessories/cleaning brush by the user while angulating.As a result, lubricant was used during the assembly of the scope leaked from the scratches.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11405795
MDR Text Key234499700
Report Number8010047-2021-03252
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403392
UDI-Public04953170403392
Combination Product (y/n)N
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2021
Initial Date FDA Received03/03/2021
Supplement Dates Manufacturer Received03/16/2021
Supplement Dates FDA Received04/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-