• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. BLAKE (R) DRAIN 10FR ROUND; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. BLAKE (R) DRAIN 10FR ROUND; CATHETER, IRRIGATION Back to Search Results
Model Number 2227
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Additional information has been requested and obtained.To date the device has not been received.If further details are received at a later date a supplemental medwatch will be sent.No further information is available.What is the lot number? no further information is available.Was another drain needed to correct the situation? no further information is available.If yes, was the new drain placed surgically during a second procedure? device return status.We regularly contact with sales rep about the device returning.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported a patient underwent an unknown orthopedic surgery on (b)(6) 2021 and a drain was used.During surgery, soon after inserting the drain to the patient, the reservoir swelled, so when checking them, there was a hole on the drain.The drain was not fixed with the suture yet, so it is unlikely that it was hurt during the operation.Further details are not provided.There were no adverse consequences to the patient.
 
Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 6/8/2021.H3 evaluation: complaint sample was received.After opening the complaint sample primary pack, it was observed that there was a small piece of drain.The complaint sample (drain) was not in original its length.The actual length of drain is 1200mm however, the size of complaint drain was 300mm.The complaint sample photograph was received for analysis but on the basis of photograph it can not be corelate the leakage due to manufacturing process.Evaluation of retain sample was not done as the lot number is not specified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Product complaint # (b)(4).Date sent to the fda: 6/8/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAKE (R) DRAIN 10FR ROUND
Type of Device
CATHETER, IRRIGATION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08876
MDR Report Key11406248
MDR Text Key234773377
Report Number2210968-2021-01937
Device Sequence Number1
Product Code GBX
UDI-Device Identifier10705031003552
UDI-Public10705031003552
Combination Product (y/n)N
PMA/PMN Number
CL I EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2227
Device Catalogue Number2227
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Initial Date Manufacturer Received 02/17/2021
Initial Date FDA Received03/03/2021
Supplement Dates Manufacturer Received05/18/2021
Supplement Dates FDA Received06/08/2021
Patient Sequence Number1
-
-