Additional information in d4, g4, h4, h6 investigation results: it was reported that the broach handle is loose and will not hold the rasp firmly.No surgical delay or injury to the patient was reported.The device, intended for use in treatment, was returned for investigation.Upon visual inspection, the reported failure mode could not be confirmed.The device shows very limited signs of use but no damage is observed on the connector site to the broach.Furthermore, the instrument was tested with gauge id-nr 0723.No deviation was detected.No additional complaint with batch a60410 was reported so far.A review of the production documentation for batch a60410 did not reveal any deviation from the standard operating procedure.Based on the performed investigations, the reported failure mode could not be confirmed.However, previous investigations demonstrated that under specific circumstances the double offset adapter 80/45 may disconnect from the rasp during backslapping.An optimized design of the device has been released in order to reduce the occurrence of this issue.This version of the device will be monitored for similar issues.
|