It was reported that the broach handle is loose and will not hold the rasp firmly.The device, intended for use in treatment, was returned for investigation.Upon visual inspection, the reported failure mode could not be confirmed.The device shows limited signs of use but no damage is observed on the connector site to the broach.Furthermore, the instrument has undergone a functional check.Instead of a premature separation problem, rather a slight physical resistance during removal from the gauge was detected, but still within the normal.No additional complaint with batch b71870 was reported so far.A review of the production documentation for the batch in scope did not reveal any deviation from the standard operating procedure.Based on the performed investigations, the reported failure mode could not be confirmed.The slight physical resistance during gage removal was still within the normal and the need for corrective action is not indicated.This version of the device will be monitored for similar issues.Should additional information become available, this complaint will be reassessed.This investigation is considered closed.
|