Model Number 3100 A |
Device Problem
Unintended Power Up (1162)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported to vyaire medical that the 3100a ventilator intermittently turning on and making a buzzing noise.As of this time, there is no information about patient involvement associated with this reported event.
|
|
Manufacturer Narrative
|
Results of investigation: vyaire field service went onsite.Found circuit would not pressurize, the column fan was very loud and the solenoid 1 is not activating when the "reset power fail" button is pressed.As a resolution the technician replaced the column fan and solenoid 1.Technician performed calibration of the unit.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Manufacturer Narrative
|
Result of investigation: failure analysis conducted visual inspection of solenoid showed the wire receptacle missing as though it was cut off and not returned with the part.At the current state the reported issue may have been related to a loose connection with the missing receptacle.Visual inspection of fan assembly found that the fan have been heavily used and after physically spinning the fan blades found to have resistance indicating the internal bearings are damaged.The fan running intermittently verifying the reported issue.As this issue arose after a 6k pm (preventive maintenance) was needed no further actions are needed at this time.Physical damage of solenoid inhibited an investigation and improper maintenance was found as the contributing factor for the fan assembly.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Search Alerts/Recalls
|