One strip from lot 449498-21 was returned for investigation where it was tested using a reference meter with a control sample, lot 47788900.Testing results: retention meter with customer strips: 2.6 inr.The obtained result was in the allowed range.No error message occurred during the investigation.Routine retention testing is performed.Retention testing data is reviewed and appropriate actions are taken as needed.The returned and the retention material meet the specifications.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
|