ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2021
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi) (b)(4).Product labeling states: "the coaguchek system uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas deviations can occur when compared to methods using other thromboplastins.However, those deviations between thromboplastins of different origin (e.G., rabbit based) are not specific to coaguchek products.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared to other laboratory methods." the customer¿s strips were requested for return.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.Test strip retention samples passed the internal inspection.Routine retention testing is performed.Retention testing data is reviewed and appropriate actions are taken as needed.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
There was a complaint of a discrepant inr result with coaguchek xs plus meter serial number (b)(4) when compared to a laboratory result using a stago analyzer.The meter result was 3.5 inr and the laboratory result was 2.66 inr.The patient's therapeutic range was asked for but not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key11417545
MDR Text Key246241064
Report Number1823260-2021-00658
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number04625315160
Device Lot Number49017711
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2021
Initial Date FDA Received03/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-