Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/05/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Associated products : item#: 00598005703; stemmed tibial component precoat size 9; lot#: 62227581.Reported event was confirmed by review of medical records.Device history record (dhr) was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review found poly would not insert and when it was removed the dovetail was not correct.Another poly was opened to complete the procedure with no further complications.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported during an initial total knee arthroplasty approximately 7 years ago, the surgeon notes he placed the poly insert and the first insert would not fit, and when it was removed, it was found that the dovetail was not correct; therefore, we used a second 12mm, it fit, and it snapped in correctly.The procedure was completed without further complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE 12 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11418477
MDR Text Key244870760
Report Number0001822565-2021-00601
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Model NumberN/A
Device Catalogue Number00596205012
Device Lot Number62632582
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/16/2021
Initial Date FDA Received03/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-