MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6379
Device Problems No Display/Image (1183); Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2021
Event Type  malfunction  
Manufacturer Narrative
The reported complaint was confirmed.The srt replaced the flow board.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The service repair technician (srt) reported that during testing of the device at the service center, the unit did not display the flow board software version, only dashes.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key11425364
MDR Text Key242209204
Report Number1828100-2021-00077
Device Sequence Number1
Product Code DWA
UDI-Device Identifier00886799000076
UDI-Public(01)00886799000076(11)110214
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/10/2021
Initial Date FDA Received03/05/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-