RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 05mar2021.
|
|
Event Description
|
The customer reported to philips that the unit was not reading tidal volume when o2 was set to 100%.The customer reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The caller advised no significant errors in the logs.The rse advised the caller to perform performance verification testing (pvt) and provided part id for flow sensor assembly.The customer (biomed) called back and reported the unit passed pvt after the flow sensor assembly was replaced but it was still not showing tidal volumes.The rse advised the customer to not use a test adapter but to use a test lung since the unit was looking for feedback.The biomed advised they ran the unit with the test lung and performed pvt and the unit successfully passed testing.The biomed later called back to report that the unit was now displaying a diagnostic code related to proximal pressure sensor auto zero failure.The customer reported that reseating the data acquisition (da) board resolved the proximal pressure sensor failure.The device passed all testing and operated within the manufacturing specifications and was returned to service.
|
|