RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event:(b)(6) 2021.Date of report: 05mar2021.
|
|
Event Description
|
It was reported that the ventilator 'shut off 'when preparing for intubation and an anesthesia mask is provided over the cannula in preparation for intubation.The ventilator was being used on a patient at the time of the reported event, however, there was no patient harm reported.Philips clinical specialist provided additional clinical education to the customer regarding the v60 alarm algorithm when using the high flow therapy (hft) option and also recommended removing the v60 hft when using the anesthesia mask.Upon a follow up, the customer reported to have 'not received a formal incident report from the hospital staff and is still researching these cases'.Although requested no further information was provided.
|
|