|
Model Number V60 PLUS |
Device Problems
Loss of Power (1475); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.08mar2021.
|
|
Event Description
|
The customer called into technical support (ts) reporting that the device¿s battery failed alarm and turns off even on ac.This event was reported to have occurred during clinical use.There was no allegation of harm associated with this event.The device was being set up when the issue occurred and that unit never came in contact with the patient.The customer used another v60.The manufacturer¿s international service technician evaluated the ventilator and confirmed the reported problem.The data acquisition rp-pcba failed after switching on the equipment.The service technician replaced the data acquisition board assembly (da pcba) and the problem was resolved.The device was put back into service.
|
|
Manufacturer Narrative
|
G4:29mar2021.B4:06apr2021.H11: b5: the customer called technical support (ts), reporting that the device displayed a machine pressure sensor auto zero failed and speaker test (pm pcba & mc pcba) error message.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4:29mar2021.B4:09apr2021.The technical support found no problem with the speaker test.The equipment passed testing.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
The data acquisition assembly (daq) was received for failure investigation.Visual inspection of the daq revealed no evidence of damage or contamination.The daq was tested and root cause is failure of daq assembly, caused by u7 drifting out of calibration.
|
|
Search Alerts/Recalls
|
|
|