It was reported that the broach handle is loose and will not hold the rasp firmly.The device, intended for use in treatment, was returned for investigation.Upon visual inspection, the reported failure mode could not be confirmed.The device shows very limited signs of use and no damage is observed on the connector site to the broach.Furthermore, the instrument has undergone a functional check.No deviation was detected.Furthermore, the device was tested within an authentic test setup to examine the reported premature separation problem.The complained connection issue between the offset adapter and the rasp could not be reproduced after 30 hard hammer strokes.Hence, the device functioned as intended.No additional complaint with batch b70751 was reported so far.A review of the production documentation for the batch in scope did not reveal any deviation from the standard operating procedure.Based on the performed investigations, the reported failure mode could not be confirmed as the device functions as intended.There is no indication that the device failed to match specification at the time of manufacturing and the need for corrective actions is not indicated.This version of the device will be monitored for similar issues.Should additional information become available, this complaint will be reassessed.This investigation is considered closed.
|