ALCON RESEARCH, LLC - HUNTINGTON MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
|
Back to Search Results |
|
Model Number ASKU |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was not returned for analysis.Product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A facility representative reported that during an intraocular lens (iol) implant procedure, the shooter did not go through, no reported patient harm.Additional information received states that a different shooter was used because it was too small for the lens and the lens did not go through cartridge.
|
|
Manufacturer Narrative
|
The lens was returned in a non company cartridge.No evaluation required.The manufacturer internal reference number is: (b)(4).
|
|