• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 8; PROST, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLY/MET/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 8; PROST, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLY/MET/POLY Back to Search Results
Model Number 74012418
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2021
Event Type  Injury  
Manufacturer Narrative
Our reference number: case-(b)(4).
 
Event Description
It was reported that during tka surgery, the journey dcf ap fem cut blk 8 sizer came apart during surgery and case ended up downsizing.The procedure was completed without delay using a smith and nephew back-up device.No other complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation, therefore the failure could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tabs of the top of the device have broken and separated a piece of the device, rendering the device inoperative.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY DCF AP FEM CUT BLK 8
Type of Device
PROST, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLY/MET/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11446534
MDR Text Key238676880
Report Number1020279-2021-01912
Device Sequence Number1
Product Code JWH
UDI-Device Identifier03596010556707
UDI-Public03596010556707
Combination Product (y/n)N
PMA/PMN Number
K111711
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74012418
Device Catalogue Number74012418
Device Lot Number13HM07184R
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/11/2021
Initial Date FDA Received03/09/2021
Supplement Dates Manufacturer Received04/07/2021
07/19/2021
Supplement Dates FDA Received04/27/2021
07/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-