LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).The customer requested deep disinfection in order to solve the reported contamination.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera.The laboratory report confirming the reported contamination has been provided to livanova.There is no known patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: through follow up communication under previous contamination cases from the same hospital, livanova learned that the device was cleaned regularly per the instruction for use and that it was placed outside the operation theater.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
MDR Report Key11453846
MDR Text Key240733458
Report Number9611109-2021-00113
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2021
Initial Date FDA Received03/10/2021
Supplement Dates Manufacturer Received04/15/2021
Supplement Dates FDA Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-