MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. DAFILON BLUE 2/0 (3) 75CM DS24; OTHER SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. DAFILON BLUE 2/0 (3) 75CM DS24; OTHER SUTURE Back to Search Results
Model Number G0935360
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2021
Event Type  malfunction  
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have received 3 closed samples and one open sample (only thread).We have tested the needle attachment strength of the closed samples received and the results fulfil the requirements of the european pharmacopoeia (ep): 2.95 kgf in average and 2.35 kgf in minimum (ep requirements: 1.12 kgf in average and 0.46 kgf in minimum).Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Final conclusion: although the results of the closed samples received fulfil the specifications of european pharmacopoeia/ b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.We regret any inconvenience this issue may have caused and thank you for your collaboration.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures,there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Event Description
It was reported an issue with dafilon suture.The client (veterinary user) reported that the thread has separated from the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 2/0 (3) 75CM DS24
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key11458221
MDR Text Key240037967
Report Number3003639970-2021-00051
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K990090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG0935360
Device Catalogue NumberG0935360
Device Lot Number619326
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/11/2021
Initial Date FDA Received03/11/2021
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-