• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-531PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-531PA; TRANSMITTER Back to Search Results
Model Number ZM-531PA
Device Problems Failure to Select Signal (1582); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2021
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the transmitter was not sending ecg(electrocardiogram) waveforms to the bedside or cns (central nursing station).When a known working transmitter was connected to the receiver and it connected and the ecg(electrocardiogram) waveforms were displayed.
 
Event Description
The biomedical engineer (bme) reported that the transmitter was not sending ecg(electrocardiogram) waveforms to the bedside or cns (central nursing station).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-531PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
MDR Report Key11464261
MDR Text Key245412851
Report Number2080783-2021-00135
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115107
UDI-Public04931921115107
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/11/2021
Distributor Facility Aware Date02/09/2021
Device Age5 MO
Event Location Hospital
Date Report to Manufacturer03/11/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-