• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD Back to Search Results
Model Number 3186
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2021
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that during the patient's permanent scs system implant surgery, the physician was unable to implant the lead due to the white cap of the stylet falling off making it impossible to use the stylet.The surgery was abandoned.
 
Manufacturer Narrative
An abandoned procedure was reported to abbott.The white cap of a lead fell off and the lead was not steerable during procedure.The results of the investigation are inconclusive as the product was not returned for analysis.As a result of this issue, a supplier corrective action report has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key11464380
MDR Text Key239186736
Report Number3006705815-2021-00902
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/12/2022
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000102098
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2021
Initial Date FDA Received03/11/2021
Supplement Dates Manufacturer Received04/21/2021
Supplement Dates FDA Received04/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
Patient Weight98
-
-