Model Number 408719
Device Problems Intermittent Capture (1080); Fracture (1260); High impedance (1291); Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Event Description
Home monitoring reported multiple pacing impedance spikes over 3000 ohms.Recordings also show noise.Advised to evaluate further and data to be presented to physician for next steps.Lead currently remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
18-apr-2023 lv capture is intermittent at high output of 7.5 v at 1.0 ms.Lead currently remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
08/jun-2023 this lead was explanted due to continued high impedance and suspicion of fracture.18-apr-2023 lv capture is intermittent at high output of 7.5 v at 1.0 ms.Lead currently remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
08/jun-2023 this lead was explanted due to continued high impedance and suspicion of fracture.18-apr-2023 lv capture is intermittent at high output of 7.5 v at 1.0 ms.Lead currently remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual, mechanical and electrical inspection.The analysis showed that the conductor coil was found fractured 35 cm distal to the connector pin, which is assumed to be the root cause of the clinical observations.Based on the characteristics, as well as the location of the damage, it is reasonable to assume that the lead was subject to excessive mechanical forces as the result of clavicular-first rib entrapment.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-85/49
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key11467098
MDR Text Key239398474
Report Number1028232-2021-01191
Device Sequence Number1
Product Code OJX
UDI-Device Identifier04035479148959
UDI-Public(01)04035479148959(17)220930
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number408719
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/09/2021
Initial Date FDA Received03/11/2021
Supplement Dates Manufacturer Received03/19/2021
04/18/2023
06/08/2023
09/07/2023
Supplement Dates FDA Received03/22/2021
04/19/2023
06/14/2023
09/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age82 YR
Patient SexFemale
-
-