• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021 11mar2021.
 
Event Description
The customer called into technical support (ts) reporting that the device is displaying error code cbit temp, i error blower.The customer reported there was no patient involvement at the time the issue was discovered.
 
Manufacturer Narrative
G4:22mar2021 b4:(b)(6) 2021 the customer reported that the device is displayed blower high temperature error.The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.The fse replaced the of motor controller board and blower to resolve reported problem.The device passed testing and is returned back to service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
The removed motor controller (mc) board and blower motor assembly were returned to the failure investigation lab (fi).A visual inspection of the mc board and blower motor assembly revealed no evidence of damage or contamination.The fi technician installed the mc board and blower motor assembly into a fi ventilator to duplicate the reported issue.The customer complaint was verified.The returned blower failed the temperature sensor test when used with the mc board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
yobana sanchez
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key11467268
MDR Text Key241381567
Report Number2031642-2021-00933
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2021
Initial Date FDA Received03/11/2021
Supplement Dates Manufacturer Received02/11/2021
11/08/2021
Supplement Dates FDA Received04/06/2021
12/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-