MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO; PORT & CATHETER, IMPLANED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO; PORT & CATHETER, IMPLANED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number UNK POWER PORT DUO
Device Problems Partial Blockage (1065); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.
 
Event Description
It was reported that sometime post port implant, the device septum was allegedly disrupted.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one powerport duo mri attached to a catheter was returned for evaluation.Visual, microscopic visual, tactile and functional evaluation were performed on the returned device.The investigation is confirmed for the reported material protrusion and identified partial blockage issues as one of the two port septa(right) was partially dislodged and upon infusing the port, leak was noted from the partially dislodged septum.Further during functional evaluation the right septum failed the mandrel test.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3, h6 (device).H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that sometime post port implant, the device septum was allegedly disrupted.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO
Type of Device
PORT & CATHETER, IMPLANED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11471680
MDR Text Key239368259
Report Number3006260740-2021-00785
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
PMA/PMN Number
K090512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK POWER PORT DUO
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2021
Initial Date Manufacturer Received 02/22/2021
Initial Date FDA Received03/12/2021
Supplement Dates Manufacturer Received08/23/2021
Supplement Dates FDA Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-