• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X MM56; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X MM56; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71332756
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: case (b)(4).
 
Event Description
It was reported that, during a thr the r3 0 deg xlpe acet lnr 36mm x mm56 cannot be anchored in the cup.Look 2-3mm out of the cup.The procedure was completed without delay using a smith-nephew back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Without the actual products involved and/or device information, our investigation cannot proceed.If these devices or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Manufacturer Narrative
The associated device was returned and evaluated.The contribution of the device to the reported event could not be corroborated.A visual inspection of the returned device did not confirm the stated failure mode.A dimensional evaluation of the returned device could not confirm the stated failure mode.The device was found to be within tolerance.A review of complaint history did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 36MM X MM56
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11480214
MDR Text Key239668756
Report Number1020279-2021-02162
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010598004
UDI-Public03596010598004
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71332756
Device Catalogue Number71332756
Device Lot Number20KM07189
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/18/2021
Initial Date FDA Received03/15/2021
Supplement Dates Manufacturer Received06/16/2021
12/29/2021
Supplement Dates FDA Received06/18/2021
01/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-