• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS) Back to Search Results
Model Number 37612
Device Problems Failure to Deliver Energy (1211); Delayed Charge Time (2586); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that there was poor coupling with the recharger and the recharger shuts off when recharging.They stated that the recharger antenna and desktop charger (dtc) are not damaged.The recharger is showing normal recharging screen but coupling boxes are not filled in.A replacement recharger was sent.The patient called back on (b)(6) 2021 and stated that they received the replacement recharger.They are now able to get 4 coupling bars filled in but they are not able to get any more than that.They have rotated the antenna dial and repositioned the antenna, but they still aren't able to get more than 4 coupling bars to fill in. the patient was redirected to their healthcare provider (hcp) to further address the issue.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the consumer reported they received a recharger replacement, but they still were unable to get any coupling bars to fill in.The patient stated that they have been charging their ins for 6-7 hours, but the ins battery level is still between 0-25% (first quartile).The patient also mentioned that the ins is off (no lightning bolt in upper left corner of the recharger display).It was reviewed that the ins battery level will need to be at least 25% before it can be turned back on.The patient was going to have their healthcare provider check their ins at their appointment next thursday or friday.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key11504827
MDR Text Key244802666
Report Number3004209178-2021-04421
Device Sequence Number1
Product Code MRU
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2020
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/15/2021
Initial Date FDA Received03/17/2021
Supplement Dates Manufacturer Received04/15/2021
Supplement Dates FDA Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age67 YR
-
-