Model Number 26280
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: a visual examination found that the stent of the device had been deployed from the delivery system.The stent was found to be damaged at more than one location.A visual inspection of the stent impression and stent cups identified no issues.The imprinted stent impression was clear on the stent holder.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile examination identified no kinks or damage along the length of the device.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 01mar2021.It was reported that stent failed to deploy.The moderately stenosed target lesion was located in the moderately tortuous and severely calcified iliac vein.A 75cm 14x90/9fr wallstent endoprosthesis stent was selected for use.However, upon flushing, the device could not deploy.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11505880
MDR Text Key240648701
Report Number2134265-2021-02778
Device Sequence Number1
Product Code MAF
UDI-Device Identifier08714729204015
UDI-Public08714729204015
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/10/2021
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0023329969
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2021
Initial Date FDA Received03/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient Weight65
-
-