• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY C CALCIUM REAGENT KIT; AZO DYE, CALCIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY C CALCIUM REAGENT KIT; AZO DYE, CALCIUM Back to Search Results
Model Number 07P5720
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2021
Event Type  malfunction  
Manufacturer Narrative
No additional patient information is available.(b)(4).Was this device serviced by a third party? no.A review of tickets was performed for reagent lot number 03169un20.The ticket search determined that there is normal complaint activity for the reagent lot.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the alinity c calcium reagent ln 7p57-20/lot number 03169un20 was identified.
 
Event Description
The customer generated falsely elevated alinity c calcium results for one patient.The customer uses the reference range 8.5-10.5 mg/dl.The following information was provided: sid:(b)(6), initial result 14.23 mg/dl, repeat 9.69 and 9.52 mg/dl.No impact to patient management was reported.
 
Manufacturer Narrative
This follow up is submitted to populate fields d8 and/or h6 with data that had previously been provided in field h10.There is no change to the content of the data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CALCIUM REAGENT KIT
Type of Device
AZO DYE, CALCIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key11510207
MDR Text Key267163801
Report Number3002809144-2021-00176
Device Sequence Number1
Product Code CJY
UDI-Device Identifier00380740132613
UDI-Public00380740132613
Combination Product (y/n)N
PMA/PMN Number
K062855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/21/2021
Device Model Number07P5720
Device Catalogue Number07P57-20
Device Lot Number03169UN20
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/25/2021
Initial Date FDA Received03/17/2021
Supplement Dates Manufacturer Received06/20/2021
Supplement Dates FDA Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY C PROCESSING MODU, 03R67-01, AC02239; ALNTY C PROCESSING MODU, 03R67-01, AC02239
-
-